0 |
No sound |
|
|
|
- |
1 |
BeepBeepBeepBeep |
|
|
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-1.wav |
Advisory |
2 |
BipBeep BipBeep BipBeep BipBeep |
End of insulin suspend. Press OK to resume basalrate delivery. |
2 sets of beeps every minute for 3 minutes. |
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-2.wav |
Advisory |
2 |
BipBeep BipBeep BipBeep BipBeep |
Auto off. Remove Pod now. |
2 sets of beeps every minute for 15 minutes. Escalates to Auto-off hazard alarm if not acknowledged within 15 minutes. |
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-2.wav |
Advisory |
2 |
BipBeep BipBeep BipBeep BipBeep |
Low reservoir. Replace Pod soon. |
2 sets of beeps every minute for 3 minutes. Repeats every 60 minutes until acknowledged. Escalates to Empty reservoir hazard alarm when empty. |
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-2.wav |
Advisory |
2 |
BipBeep BipBeep BipBeep BipBeep |
Pod expiration advisory. Change Pod now. |
2 sets of beeps repeat periodically. Repeats every 15 minutes during the last hour of the Pod' s life and is independent of the Expiration notification that you can set. Escalates to Pod expired hazard alarm. |
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-2.wav |
Acknowledge |
3 |
BipBip |
Pod start/Pod end |
|
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-3.wav |
- |
4 |
Beep |
Pod in Suspended mode |
Beeps every 15 minutes |
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-4.wav |
- |
5 |
BeepBeepBeep |
Suspend in progress |
|
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-5.wav |
Acknowledge |
6 |
Beeeeeep |
Cancel Temp Basal |
|
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-6.wav |
- |
7 |
BipBipBip BipBipBip |
Timer limit |
|
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-7.wav |
- |
8 |
BeeepBeeep |
|
|
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-8.wav |
Hazardous |
|
Beeeee...eeep |
Replace pod |
Continuous tone |
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-F.wav |
Hazardous |
|
Beeeee...eeep |
Empty reservoir. Insulin delivery stopped. |
Continuous tone |
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-F.wav |
Hazardous |
|
Beeeee...eeep |
Pod expired. Insulin delivery stopped. Change Pod now. |
Continuous tone |
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-F.wav |
Hazardous |
|
Beeeee...eeep |
Occlusion detected. Insulin delivery stopped. Change Pod now. |
Continuous tone |
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-F.wav |
Hazardous |
|
Beeeee...eeep |
Pod Error. Insulin delivery stopped. Change Pod now. |
Continuous tone |
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-F.wav |
Hazardous |
|
Beeeee...eeep |
System Error. Remove Pod now. |
Continuous tone |
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-F.wav |
Hazardous |
|
Beeeee...eeep |
Pod deactivation required. Press "Ok" to deactivate. This alarm occurs after a PDM error occurs that requires the user to reset the date and time. |
Continuous tone |
https://github.com/openaps/openomni/blob/master/audio_recordings/1f-nibble-F.wav |