DICOM SR Exploration of relevant templates - QIICR/ProjectIssuesAndWiki GitHub Wiki
Of historical interest - specific templates have now been identified and standardized and are summarized at DICOM CPs and Supplements for QIICR
Identify previous work in DICOM SR templates (existing templates that are/are not part of the standard) that is relevant in the context of encoding analysis results for DBPs.
- Explore the capabilities of Encoding of Oncology Clinical Trial Read Results in DICOM SR (QIBA VolCT, contributed by David Clunie)
- Consider templates referenced by David in his presentation (TID 1410 Planar ROI measurements, TID 1420 Volumetric ROI measurements (defined in DICOM CP 1112 ftp://medical.nema.org/medical/dicom/final/cp1112_ft.pdf)
Google doc to abbreviate main concepts of the QIBA Clinical trial results template (in progress)
Observations:
- concepts of reader, lesion, time point, subject are present only in the template that is not part of the standard
- algorithm parameters description is addressed at a very rudimentary level: some of the parameters encoded are very specific to the algorithm used (e.g., TID RPH_0007b: Sub-Regions and Measurements), while in other places these are completely unstructured (TID RPH_0002b: Device Observer Identifying Attribute)
- parameter maps are not part of any of the existing templates (need to confirm)