Spectrophotometer - MetabolicEngineeringGroupCBMA/MetabolicEngineeringGroupCBMA.github.io GitHub Wiki
Procedure
-
Turn on the spectrophotometer GENESYS20
- Allow the instrument to warm up for at least 10 minutes if required by the manufacturer.
-
Set wavelength
- Adjust the spectrophotometer to 640 nm.
-
Prepare the blank
-
Fill a clean cuvette with 1 mL (or the required volume) of the same sterile medium used to grow the culture.
-
Wipe the outside of the cuvette with a lint-free tissue to remove fingerprints or droplets.
-
-
Blank the instrument
-
Insert the blank cuvette into the holder, aligning the transparent sides with the light path.
-
Close the lid and press βBlankβ or βZeroβ.
-
Wait until the display reads 0.000 absorbance (or 100% transmittance).
-
-
Measure the sample
-
Mix the microbial culture gently by inversion or pipetting (do not vortex to avoid cell lysis or bubbles).
-
Transfer 1 mL of the culture into a clean cuvette.
-
Wipe the cuvette and place it in the same orientation as the blank.
-
Record the ODβββ value.
-
-
Dilute if necessary
-
If the ODβββ value exceeds 0.5, dilute the sample (e.g., 1:10 in fresh medium), mix, and remeasure.
-
Multiply the measured ODβββ by the dilution factor to obtain the true ODβββ.
-
-
Clean up
-
Rinse cuvettes thoroughly with distilled water and let them air-dry or store according to lab policy.
-
Turn off the spectrophotometer if not in continuous use.
-
Notes
-
Always use the same type of cuvette (plastic or glass) for blank and samples.
-
Avoid bubbles and residual dropletsβthey can alter the reading.
-
Record all values in a lab notebook or data sheet, including dilution factors and time points.
1. The origin of ODβββ
-
ODβββ (600 nm) became the standard wavelength for measuring bacterial growth, particularly E. coli, because:
-
Early photoelectric colorimeters and spectrophotometers (mid-20th century, e.g., Klett-Summerson colorimeter) used filters centered around 600 nm (βKlett filter #66β was 600 Β± 20 nm).
-
At 600 nm, visible light scattering by bacterial cells is high, but absorption by medium components (especially proteins and riboflavin) is minimal.
-
It provided a convenient linear correlation between cell density and light scattering up to OD β 0.8β1.0.
-
Consequently, ODβββ was widely adopted in bacterial growth protocols, and it became embedded in textbooks and instrument presets.
-
2. Why ODβββ (or ODβββββββ) is sometimes used
-
Yeast (e.g., Saccharomyces cerevisiae) cells are larger and more refractile than bacteria; they scatter light differently.
-
Early yeast researchers found that measurements at slightly longer wavelengths (620β640 nm):
-
Reduced multiple scattering (less overestimation at high cell densities).
-
Improved linearity between OD and actual cell dry weight or concentration.
-
Minimized interference from colored media or metabolic pigments (like flavins or residual YPD coloration).
-
-
Some instruments (especially older spectrophotometers in yeast labs) came equipped with a 640 nm or 620 nm filter, not 600 nm, simply because they were originally designed for turbidity or colorimetric assays (e.g., McFarland standards, enzymatic reactions).
-
As a result, ODβββ became the norm in many yeast laboratories β particularly in Europe (e.g., Scandinavian and German microbiology traditions).
3. Practical differences
-
Both wavelengths measure scattered light, not true absorption.
-
The actual difference in measured OD between 600 and 640 nm is typically small (β 5β10%), but not negligible if one is comparing across studies.
-
Empirically, some labs use correction factors:
[
\text{OD}{600} β 1.1 Γ \text{OD}{640}
]
(though this varies depending on the spectrophotometer and culture conditions).
4. In the literature
-
E. coli and most bacteria β ODβββ (historical Klett tradition, standard in molecular biology).
-
S. cerevisiae and other yeasts β ODβββ or ODβββ (to improve linearity with biomass and because of larger cell size/scattering behavior).
-
Microalgae and colored cells β even longer wavelengths (e.g., 680 nm) to minimize chlorophyll absorption overlap.
5. Bottom line
The difference is not biochemical, but optical and historical:
ODβββ comes from bacterial tradition (Klett filter #66).
ODβββ emerged from yeast work to improve scattering linearity and reduce medium interference.
Both are valid, as long as:
-
You consistently use one wavelength within your lab.
-
You report it explicitly (e.g., βCell density was monitored at ODββββ) when publishing.