ZOOM Instrument Details - ISISComputingGroup/ibex_developers_manual GitHub Wiki
This page collects information that will be useful for the implementation of the new control system on ZOOM.
Background & Timeline
The ZOOM web page describes the background to the instrument. Additional material is also available on the ZOOM sharepoint. ZOOM is expected to have 36 motion axes.
The ZOOM shutter is due to open in February 2017 (see Zoom plan 28th May 2016).
Control System
ZOOM will use the IBEX control system.
ZOOM Equipment
The equipment listed below is to be used on ZOOM. Please add new information (e.g. new items of equipment, new notes, information about drivers, etc. to this table).
Manufacturer | Model | Type | Connection | Driver | Notes |
---|---|---|---|---|---|
ISIS | Fast Shutter | OMRON | see Fast Shutter note | ||
ISIS | Double Disc Chopper | CHOPPER | Ethernet/.NET | #170 | see ISIS Double Disk Chopper note |
JAWS | #179, #180 | see Jaws note | |||
Polariser, guide & collimation unit (PGC) | #1233 | see Polariser, Guide & Collimator note | |||
JAWS | #179, #180 | ||||
ISIS | Spin Flipper | #1234 | see SpinFlipper note | ||
JAWS | #179, #180 | ||||
Alignment Laser | see Alignment Laser note | ||||
JAWS | #179, #180 | ||||
JJ XRAY | 7-axis Sample Stack Manual | Sample Stack | #304, #1238 | see JJ X-Ray note | |
Translation (across beam) | Sample Stack | #1238 | see SampleStack note | ||
ISIS | Translation (along beam) | Sample Stack | #1238 | Driven by Beckhoff | |
ISIS | Two Tier - possibly | Sample Changer | see SampleChanger note | ||
Pneumatic Shutter | OMRON | see Pneumatic Shutter note | |||
ISIS | Tank Translation | see Tank Translation noteDriven by Beckhoff | |||
In/Out Monitor | #267 | see Monitor note | |||
JAWS | #179, #180 | ||||
Detector Motion System (DMS) | #1236 | see Detector Motion System (DMS) note | |||
ISIS | DAE (2 or 3?) | Detector Electronics | Ethernet | #219 | see DAE note |
CAEN | HT | Detector HV Control | Ethernet | #214 | see CAEN HT note |
OMRON | PLC | Ethernet | FINS, #217 | see OMRON PLC note | |
Galil | DMC2280 | Motion Controller | Ethernet | EPICS | see Galil note |
Beckhoff | Motion Controller | TBC | see Beckhoff note | ||
Pfeiffer | TPG300 | ISIS VACUUM SYSTEM | RS232 | #218 | see Pfeiffer note |
Julabo | TBC | Water Bath | RS232 | #187, #188 | see Julabo note |
PSU | TBC | see Power Supply Unit note | |||
Oxford Instruments | ITC503 | Temp Controller | RS232 | see Oxford Instruments Temperature Controller note | |
Oxford Instruments | ILM200 series | Helium Level Meter | RS232 | EPICS | see Oxford Instruments Helium Level Meter note |
Eurotherm | All models at ISIS | Temp Controller | RS232 | EPICS | see Eurotherm note |
Eurotherm | Nanodac | Temp Controller | Ethernet | see Eurotherm note | |
Thurlby | EX355P | DC Power Supply | RS232 | #156 | see Thurlby note |
KEPCO | TBC | Bi-Polar Power Supply | RS232 | #189 | see KEPCO note |
Tektronix | MSO 4104B | Oscilloscope | Ethernet | #234 | see Tektronix note |
Tektronix | AFG 3022B | Function Generator | Ethernet | #237 | see Tektronix note |
PI | Rotation Stage | Nano-positioning Rotation Stage | RS232 | see Physik Instrument note | |
Bio-Logic | Stopped Flow Cell | RS232 | see BioLogic note | ||
17T Cryomag | Super conducting magnet | RS232 | |||
Oxford Instruments | 7.5T Magnet | Super conducting magnet | RS232 | see 7.5T Magnet note | |
Scientific Magnetics | 3 Axis Super conducting magnet & VTI | RS232 | see Scientific Magnetics note | ||
Lakeshore | All models at ISIS | TBC | see Lakeshore note | ||
Syringe Pumps | RS232 | see Syringe Pumps note | |||
see Hitachi note | |||||
TS-Haake | Water Bath | RS232 | see Thermo Scientific (Haake) note | ||
Huber | Sample Stack | RS232 | see Huber note | ||
Thar | Pressure Cell | TBC | see Thar note | ||
Goudsmit | 1T Magnet | Magnet | TBC | see Goudsmit note | |
Anton Paar | Rheometer | see Pfeiffer note | |||
ISIS | Couette Cell | see ISIS Couette Cell note | |||
McLennan | Motion Controller | RS232 | see McLennan note | ||
T-Jump Cell | RS232 | see T-JumpCell note | |||
Dynamic Light Scattering | DLS | TBC | see Dynamic Light Scattering (DLS) note | ||
Raman Spectroscopy | TBC | see Raman Spectroscopy note | |||
Linkam | RS232 | see Linkam note | |||
Keithley | All models in use at ISIS | RS232, USB | see Keithley note | ||
Closed Cycle Refrigerator | via Eurotherm? | see CCR note |
Note: Fast Shutter
Location: Front End Due: Day 1 As used on: SANS2D
Note: ISIS Double Disk Chopper
ISIS Mk3 Chopper Location: Front End Due: Day 1 As used on: SANS2D
Note: Jaws
Location: Front End Vacuum Tank Due: Day 1 As used on: Many instruments Testing Summer 2015
Note: Polariser, Guide & Collimator
Location: Front End Vacuum Tank Due: Day 1 1 horizontal motion axis controlling four 4m evacuated tubes
Note: SpinFlipper
Location: Fixed Collimation Due: Day 1 4m evacuated tube & Spin-flipper unit Spin-flipper will be controlled by 2 Kepco Power Supplies
Note: Alignment Laser
Location: Blockhouse Due: Day 1 Single in/out axis No control of the alignment laser is required, other than the ability to move it in/out of the beam. It can treated like a monitor.
Note: JJ X-Ray 7-axis Sample Stack
Location: Blockhouse Due: Day 1 7-axis Sample Stack Manual See also LARMOR
Note: SampleStack
Location: Blockhouse Due: Day 1 The JJ X-Ray sample stack will sit on rails to allow the entire sample stack to be translated back & forth along the beam. The rails are not perfectly straight, so there needs to be an axis allowing motion perpendicular to the rails to correct for any slight misalignment in the rails. In other words, there are two axes for wholesale motion of the sample stack (in addition to the 7 internal axes of the sample stack itself).
- Across beam (i.e. perpendicular to the rails): Alignment stage for reproducibility.
- Along beam (i.e. along the rails): Monitoring only, no control required. See ZOOM_Sample_Stack_Schematic
- The axis setup is described here
Note: Pneumatic Shutter
Location: Blockhouse Due: Day 1 As used on: SANS2D
Note: Tank Translation
Location: Blockhouse Due: Day 1 The tank will be moved by a motor connected to a Beckhoff controller. Monitoring only, no control required.
Note: Monitor
Location: Tank Due: Day 1 In/Out positioning - c.f. LARMOR, IMAT,others
Note: Detector Motion System (DMS)
Location: Tank Due: Day 1
7 axes consisting of:
- Baffle
- Detector translation
- Strip Beam Stop
- 2x Disk Beam stop - x and y movement (total: 4 axes)
Note: DAE
Location: Screened Room Due: Day 1 As used on: Many instruments
Note: CAEN HT
Location: Screened Room As used on: LARMOR
Note: OMRON PLC
Location: TBC As used on: LARMOR, IMAT A full memory map is available There are up to three functions for the PLC(s) - exactly which are to be included in the computing control is TBC. Current expectation is that the PLC will control:
- "Fast Shutter" (located at the front end of the beamline). SANS2D has a similar shutter.
- "Pneumatic Shutter" (located in the blockhouse). SANS2D has a similar shutter.
- Pfeiffer TPG 300 pressure gauges
Note: Galil
Location: Blockhouse x 1, Mezzanine x 4 Due: Day 1 As used on: Many instruments There are at present 39 axes to be controlled via Galil
Note: Beckhoff
Location: TBC As used on: IMAT The Beckhoff controller will control two axes:
- the motion of the JJ X-ray sample stack along the rails see SampleStack
- the motion of the tank (enclosing the detectors & detector motion system).
IBEX is NOT required to actually control these two axes (that will be done independently). IBEX is required to monitor the locations reported by the Beckhoff system - a read only driver will be required.
Note: Pfeiffer
Location: TBC - alongside PLCs Due: Day 1 As used on: LARMOR, IMAT
Note: Julabo
Location: TBC Due: Day 1 As used on: LARMOR
Note: Power Supply Unit
Location: TBC Due: Day 1 This is to control a magnet which will be linked to the interlocks, but there are no details at present
Note: Oxford Instruments Temperature Controller
Model: ITC503 Location: TBC As used on: LARMOR
Note: Oxford Instruments Helium Level Meter
Model: ILM200 series As used on: LARMOR
Note: Eurotherm
Model: all models at ISIS Due: Day 1 Location: TBC As used on: LARMOR Note: ZOOM may also use a Eurotherm Nanodac to monitor detector temperatures.
Note: Thurlby
Location: TBC As used on: LARMOR
Note: Pfeiffer
Location: TBC As used on: LARMOR
Note: KEPCO
Location: TBC As used on: LARMOR
Note: Tektronix
- Model: Oscilloscope
- Model: Function Generator
Location: TBC As used on: LARMOR
Note: SampleChanger
Location: Sample Area Due: Day 1 Same as used on: SANS2D/LARMOR, translations done via sample stack
Note: Physik Instrument
Model: Nano-positioning Rotation Stage Location: TBC PI Software, EPICS driver As used on: LARMOR
Note: BioLogic
Model: Stopped Flow Cell Location: TBC As used on: LARMOR
Note: Scientific Magnetics
Model: 3 Axis Super conducting magnet & VTI Location: TBC As used on: LARMOR
Note: Lakeshore
Model: all models at ISIS Location: TBC As used on: LARMOR
Note: Pfeiffer
Model: all models at ISIS Location: TBC As used on: LARMOR
Note: Hitachi
Hitachi L-7100 HPLC pump (no longer supported at ISIS). Can't find L-7100 on Hitachi web-site (may be obsolete). Check existing SECI VI for logic and manual: both VI and manual are located here: C:\LabVIEW Modules\Drivers\Hitachi L-7100
.
- Update (27-06-2019): Hitachi pumps are no longer used. No longer any need to support them. See #3780.
Note: Thermo Scientific (Haake)
Model: Water bath Location: TBC As used on: LARMOR
Note: HUBER
Model: Sample Stack As used on: SANS2D Location: TBC
Note: Thar
Model: Pressure Cell Location: TBC As used on: SANS2D, LOQ
Note: Goudsmit
Model: 1T Magnet Location: TBC As used on: SANS2D, LOQ
Note: Pfeiffer
Model: Rheometer Location: TBC As used on: SANS2D Currently controlled from a separate PC
Note: ISIS Couette Cell
Location: TBC As used on LOQ
Note: McLennan
Location: TBC Due: Day 1 As used on: Many instruments
Note: T-JumpCell
Location: TBC As used on: LOQ
Note: Dynamic Light Scattering (DLS)
Location: TBC As used on: SANS2D There is an interaction with a LabVIEW function that ought to be reproduced.
Note: Raman Spectroscopy
Location: TBC As used on: INTER This may not have an existing interface, and may not need one
Note: Linkam
Location: TBC There are VIs for this
Note: Keithley
Model: all models in use at ISIS Location: TBC There are VIs for this
Note: CCR
Model: TBC Location: TBC Details of Closed-Cycle Refrigerators at ISIS.
Note: 7.5T Magnet
Model: 7.5T Magnet