This page collects information that will be useful for the implementation of the IBEX control system on LET.
Background & Timeline
LET is a cold neutron multi-chopper spectrometer on TS2 at ISIS. The LET web page describes the background to the instrument.
Control System
It is proposed to migrate from the SECI control system to the IBEX control system in time for Cycle 2018/04. LET does currently use IBEX to control its SKF choppers (but that's all).
LET Equipment
The equipment listed below is used on LET. Please add new information (e.g. new items of equipment, new notes, information about drivers, etc. to this table).
Manufacturer | Model | Type | Connection | Driver | Notes |
---|---|---|---|---|---|
ISIS | DAE 2 | Detector Electronics | Ethernet | see DAE note | |
SKF | MB4150-G5 | Disk Chopper | Ethernet | see High-Speed SKF Disk Chopper note | |
Astrium | TBD | Disk Chopper | TBD | .NET | see High-Speed Astrium Disk Chopper note |
ISIS | Mk3 | Disk Chopper | Ethernet | .NET | see ISIS Chopper note |
N/A | N/A | Jaws | see Jaws note | ||
Pfeiffer | TPG300 | ISIS Vacuum System | RS232 | #216 | see Pfeiffer note |
Eurotherm | Temperature Controller | All models at ISIS | RS-232 | EPICS | see Eurotherm note |
ISIS | Orange Cryostat | Cryogenic System | RS-232 | see Orange Cryostat note | |
Oxford Instruments | Heliox | Dilution Fridge | RS-232 | #3739 | see Oxford Instruments note |
Oxford Instruments | Triton | Dilution Fridge | RS-232 | #2744 | see Oxford Instruments note |
ISIS | ??? | Mezei Flipper | #3738 | see Mezei Flipper Notes | |
ISIS | Furnace | High temperature furnace | RS-232 | see Furnace note | |
Oxford Instruments | 9T | Magnet | ??? | #2765 | see 9T Magnet note |
Oxford Instruments | 14T | Magnet | ??? | see 14T Magnet note | |
Lakeshore | 340 | Temperature Controller | RS-232 | EPICS | see Lakeshore note |
ISIS | ORC | Collimator | via Galil | #2535 | see LET Notes |
ISIS | McLennan | Sample Changer | via Galil | see LET Notes | |
ISIS | McLennan | Rotation Stage | via McLennan | see LET Notes | |
ISIS | McLennan-Newport | Rotation Stage | via McLennan | see LET Notes |
Note: DAE
Main Detector banks + several fixed monitors.
Note: High-Speed SKF Disk Chopper
LET currently has 2 high-speed SKF choppers (same as IMAT). These are currently controlled using a mini-IBEX system.
- SKF G5 choppers also used on IMAT (see #1907).
- SKF MB4150-G5 Chopper
- MBScope software for SKF magnetic bearing choppers.
Note: High-Speed Astrium Disk Chopper
LET currently has 2 high-speed Astrium choppers. These are due to be replaced by 2 new SKF G5 choppers within 9-12 months (from September 2017). Latest news: The Astrium choppers will be replaced in early 2020 (one might even be replaced as early as August 2019)., but more likely is chopper 5 in January/February shutdown and chopper 1 in the long shutdown.
The Astrium choppers are currently controlled by 2 VMs running on NDHLET (one VM per chopper). The VMs communicate with the choppers, via a MOXA, using a .NET connection.
Note: ISIS Disk Chopper
LET currently has 1 Mk3 ISIS chopper.
Note: Jaws
LET has no jaws.
Note: Pfeiffer
Model TPG 300, used for vacuum system read-back.
Note: Eurotherm
Eurotherms in use on LET. Used to control temperature of all top-loading CCRs, Orange cryostats, heaters and furnaces.
Note: Lakeshore
NOTE: this entry added 07-02-2019. LET uses a Lakeshore Model 340 cryogenic temperature controller.
- Model 340.
- According to the Lakeshore web-site, the Model 340 is now obsolete. It has been replaced by the models 336 and 350.
Note: Orange Cryostat
Controlled via Eurotherm.
Note: Oxford Instruments
- Dilution fridges: models. LET uses both Triton and Heliox dilution fridges
- Triton
- Triton LabVIEW Driver : C:\LabVIEW Modules\Drivers\Triton
- Triton Documentation : C:\LabVIEW Modules\Drivers\Triton\Documentation
- See also #2915
- Heliox
- The Heliox dilution fridge is currently (November 2018) broken and is being repaired at Oxford Instruments. There is no ETA on when it will be fixed. There are no experiments scheduled to use the Heliox dilution fridge. Obviously, that will change when the Heliox is repaired.
- Triton
Note: ISIS 9T Chopper Magnet
ISIS 9T Chopper Magnet See also:IBEX Support for 9T chopper magnet.
Note: ISIS 14T Magnet
LET has, in the past, used the ISIS 14T Magnet. This magnet is currently being repaired and there is no ETA on when it might return to service.
Note: Mezei Neutron Spin Flipper
The Mezei Neutron Spin Flipper is a collection of devices (power supplies, signal generators & control PC) housed in a mobile rack. The control PC runs a Python script to control the power supplies & signal generators. The Python script accepts a set of simple commands. The Mezei Neutron Spin Flipper is, at the time of writing, used on LET and POLREF. In future, it may be used on other instruments too.
There is a VI to control the PSU, located here: C:\LabVIEW Modules\Instruments\POLREF\POLREF Flipper
. It was originally developed by Gareth Howells, but has subsequently been modified by students. Consult with Goran Nilsson & Andrew Caruana for more details.
N.B.: LET plans to use this device in Cycle 2018/04.
Note: ISIS High temperature furnace
Controlled via Eurotherm. Furnaces are used only very rarely on LET.
LET Notes
LET has the following specialist systems:
- LET specific oscillating radial collimator. Controlled by special program downloaded into Galil.
- See 2535
- LET sample changer
- A sample changer has not often been used on LET. From Cycle 2018/04 LET will be using a sample changer of the same design as the MERLIN sample changer (the two devices should be interchangeable).
- LET rotation stage
- LET has two rotation stages;
- A McLennan rotation stage (same as MERLIN, used with CCRs & Orange Cryostats))
- A McLennan-Newport rotation stage (used with the 9T Magnet)
LET SECI Configs
LET has a number of SECI configurations. Equivalent IBEX configs for the following combinations of equipment are required:
- Eurotherm (for Cryostat or CCR) + rotation stage
- Eurotherm (for Cryostat or CCR) + ORC + rotation stage
- Eurotherm (for Cryostat or CCR) + dilution fridge (Triton) + rotation stage
- Eurotherm (for Cryostat or CCR) + dilution fridge (Triton)+ ORC + rotation stage
- Eurotherm (for Cryostat or CCR) + dilution fridge (Heliox) + rotation stage
- Eurotherm (for Cryostat or CCR) + dilution fridge (Heliox)+ ORC + rotation stage
- 9T Magnet + McLennan-Newport rotation stage
LET Genie Scripts
LET has a number of instrument specific Genie scripts. The key scripts to be migrated to genie_python are:
Script | Purpose | Notes |
---|---|---|
park_choppers |
It parks the choppers, of course | Needed because LET has 4 choppers |
set_ei5 |
Sets chopper speed/phase | This is the LET version of set_ei . It is not the same as set_ei used on other excitations instruments. It would be helpful if set_ei5 could be extended to better support vetoes and wiring tables. Ignore the older versions set_ei1 - set_ei4 . |
mono |
On LET user scripts are usually created by modifying an old user script. Create a template genie_python script (based on an old user script) which can be used in future.