Alarm Information - sr320/LabDocs GitHub Wiki
Description of active alarms in the lab, how to configure, and what phone calls mean.
Alarm: Sensaphone Model 1104
User Manual here
Sensaphone Phone Number: 1-206-685-7806
This allows for you to call the alarm and receive updates on connected alarms, as well as listen to what is happening inside of the room for 15 seconds.
Alarm call list:
- 1: 206-866-5141 : Steven
- 2: Sam
- 3: 206-914-3735 :
- 4: 206-685-3273 : Sam
Alarms:
- 1: Temperature sensor for ambient temperature inside of room.
- 2: Normally closed trigger for -80C VWR freezer (Roberts Lab)
- 3: Normally closed trigger for -80C Revco freezer (formerly Horner-Devine Lab)
- 4: Not hooked to anything.
Temperature alarm sequence:
Power outage alarm sequence:
When the power goes off, Sensaphone waits 10 minutes, then issues an alarm status. It then waits 5 additional minutes to begin calling numbers on the list. There is a 1 minute pause between phone number calls.
To add/change phone numbers (The Sensaphone can hold 4 phone numbers):
- Press Set
- Press Phone Number
- Press Number on number pad corresponding to number you would like to change
- Enter Phone number on number pad
- Press Enter
- Sensaphone should respond with "Enter" if number updated correctly.
VWR alarm set points: -60* High, -90* Low.
Revco alarm set point:
Alarm: Avtech
Alarms:
- 1: FTR-209 Freezer [High Alarm: 5, Low Alarm: -50]
- 2: FTR-209 Refrigerator [High Alarm: 15, Low Alarm: -5]
- 3: FTR-213 Freezer [High Alarm: 5, Low Alarm: -50]
- 4: FTR-213 Refrigerator [High Alarm: 15, Low Alarm: -5]
Current numbers programmed to be called in the event of an alarm state:
- 1: 206-866-5141 : Steven
- 2: 206-685-3273 : Sam