Run the Demonstration (AD738x) - ArrowElectronics/data-storm-daq GitHub Wiki
The AD738x requires a Differential signal from a signal source. AD738x accepts input voltage from GND to VCC centered around VCM=(GND-Vref)/2, and differential input full voltage swing equal to Vref, above which ADC conversion will saturate. The AD738x, in its default configuration, is setup with Vcc equal to +3.3V. Vref can be set to +2.5V / +3.3v from onboard supply(jumper LK4) or through internal Vref. Here Vref set to +3.3V external so AD738x requires the input signal to be centered around VCM equal to (Vref/2 i.e +3.3V/2) which can be externally given through VCM input on EVAL-AD738x-FMCZ. Connect the Differential signal source to AinA+ and AinA- input of AD738x. Give external noise free variable VCM supply to VCM input & adjust it in range 1V to 2V for better output in FFT plot. A single ended signal generator with 2 outputs can be used to generate the differential signals driving the AD738x inputs by setting second output to track first output in inverted mode on a signal generator.
Power up the unit using this hardware configuration. Wait approximately 60 seconds for the unit to be fully functional.
IIO Oscilloscope will be automatically launched on the system desktop. No configuration is required.
Note: The snapshot demo image are based on AD7380, But similarly it can be applicable for other AD738x platforms.
Select a Plot Channel to be displayed in IIO Oscilloscope. Click on the check box in the Plot Channel window adjacent to voltage0. Right click on ad7380 in the Plot Channels window. Select voltage0 and rising edge as the trigger parameters.
Select an appropriate number of Time Domain samples in the Plot Type window. Press the Capture / Stop button to begin the capture process.
Change the capture mode to Frequency Domain. Select an FFT Size and Average. Press the Capture / Stop button to begin the capture process.
Right click in the IIO Oscilloscope plot area. Select Single Tone markers. Press the Capture / Stop button to begin the capture process with markers.
Note the information in the Marker window. This capture shows a signal source of 10 khz with harmonic components.
-
Power down the target. Launch a Root Terminal. Type the following
$ poweroff
-
Wait 30 seconds and then you can safely power down the DataStorm DAQ
A wired ethernet point-to-point connection between the host PC and the embedded target is required. Use this link for instructions on how to assign a static IP address to the host Ethernet adapter. Set the following values
-
IP address : 192.168.0.1
-
Subnet mask : 255.255.255.0
-
No entries required for the other fields
- A wired Micro USB serial port connection between the host PC and the embedded target is required
- Launch a linux terminal (like Tera Term VT or Putty) and connect using serial port
- Select 115200 baud
- Select the appropriate target COM port
Type the following at the terminal prompt to set the target IP address
- $ ifconfig eth0 192.168.0.2 up
Note: IIO Oscilloscope creates a .osc_profile.ini file in the Users/UserName/Appdata/Local directory. The next time IIO Oscilloscope is opened, the existence of this file delays connection to the target. It is recommended to delete the .osc_profile.ini (if it exists) prior to launching IIO Oscilloscope.
- Select the Discoverable/Scan option and click the Refresh button..
- A successful connection results in discovery of the IIO devices. Press OK to continue.
Note: The snapshot demo image are based on AD7380, But similarly it can be applicable for other AD738x platforms.
Select a Plot Channel to be displayed in IIO Oscilloscope. Click on the check box in the Plot Channel window adjacent to voltage0. Right click on ad7380 in the Plot Channels window. Select voltage0 and rising edge as the trigger parameters.
Press the Capture / Stop button to begin the capture process.
Change the capture mode to Frequency Domain. Select an FFT Size and Average. Press the Capture / Stop button to begin the capture process.
Right click in the IIO Oscilloscope plot area. Select Single Tone markers. Press the Capture / Stop button to begin the capture process with markers.
Note the information in the Marker window. This capture shows a signal source of 10 khz with harmonic components.
-
Power down the target. Type the following at the terminal prompt
$ poweroff
-
Wait until the linux terminal shows
$ reboot : System Halted
-
Close out IIO Oscilloscope and all associated windows
Prev- Install Required Software